A B U S and H H U S

Purpose: The purpose of this study was to prospectively evaluate the interobserver agreement
on lesion characterization and the final assessment of automated whole breast ultrasonography
(ABUS) images.
Methods: Between March and August 2012, 172 women underwent bilateral ABUS before
biopsy guided by handheld ultrasonography (HHUS) and mammography. A total of 206 breast
lesions were confirmed histopathologically by biopsy. Three-dimensional volume data from ABUS
scans were analyzed by two radiologists without the knowledge of HHUS results or patient
clinical information. The two readers described the type, shape, orientation, margin, echogenicity,
posterior acoustic features, and categorization of the final assessment of detected breast lesions.
Kappa statistics were used to analyze the described characteristics of the breast lesions detected
by both of the two readers.
Results: Of the 206 histopathologically confirmed lesions, reader 1 detected 166 lesions and
reader 2 detected 150 lesions. A total of 145 lesions were detected by both readers using ABUS
images. There was substantial agreement on shape (κ=0.707), and moderate agreement on type,
margin, mass orientation, echogenicity, and posterior acoustic features (κ=0.592, 0.438, 0.472,
0.524, and 0.541, respectively). Breast Imaging Reporting and Data System final assessment
values yielded a kappa value of 0.3971 when category subdivisions 4A, 4B, and 4C were
included. With respect to the C2, C3, C4, and C5 categories, the interobserver agreement was
moderate (κ=0.505).




Conclusion: ABUS is a promising diagnostic tool with a good interobserver agreement,
comparable to that of HHUS.
Keywords: Breast; Ultrasonography; Observer variation

Thứ Năm, 13 tháng 11, 2014

C A Dx MODELS in BREAST CANCER SCREENING



US CADx models
Ultrasound imaging is an adjunct to diagnostic mammography, where CADx models could be
used for improving diagnostic accuracy. CADx models developed for US scans date back to
late 1990s. In this section, we review studies that apply CADx systems to breast sonography
or US-mammography combination in distinguishing malignant from benign lesions. A
summary list for the primary US CADx models is presented in TABLE 2.
Giger et al. classified malignant lesions in a database of 184 digitized US images [47]. Biopsy,
cyst aspiration or image interpretation alone were used to confirm benign lesions, whereas
malignancy was proven at biopsy. The authors utilized an LDA model to differentiate between
benign and malignant lesions using five computer-extracted features based on lesion shape and
margin, texture, and posterior acoustic attenuation (two features). ROC analysis yielded AUCs
of 0.94 for the entire database and 0.87 for the database that only included biopsy- and cyst-
proven cases. The authors concluded that their analysis demonstrated that computerized
analysis could improve the specificity of breast sonography.
Chen et al. developed an ANN to classify malignancies on US images [48]. A physician
manually selected sub-images corresponding to a suspicious tumor region followed by
computerized analysis of intensity variation and texture information. Texture correlation between neighboring pixels was used as the input to the ANN. The training and testing dataset
included 140 biopsy-proven breast tumors (52 malignant). The performance was assessed by
AUC, sensitivity and specificity metrics, which yielded an AUC of 0.956 with 98% sensitivity
and 93% specificity at a threshold level of 0.2. The authors concluded that their CADx model
was useful in distinguishing benign and malignant cases, yet also noted that larger datasets
could be used to improve the performance.
Later, Chen et al. improved on a previous study [48] and devised an ANN model composed
of three components: feature extraction, feature selection, and classification of benign and
malignant lesions [49]. The study used two sets of biopsy-proven lesions; the first set with 160
digitally stored lesions (69 malignant) and the second set with 111 lesions (71 malignant) in
hard-copy images that were obtained with the same US system. Hard-copy images were
digitized using film scanners. Seven morphologic features were extracted from each lesion
using an image-processing algorithm. Given the classifier, forward stepwise regression was
employed to define the best performing features. These features were used as inputs to a two-
layer feed-forward ANN. For the first set, the ANN achieved an AUC of 0.952, 90.6%
sensitivity and 86.6% specificity. For the second set, the ANN achieved an AUC of 0.982,
96.7% sensitivity and 97.2% specificity. The ANN model trained on each dataset was
demonstrated to be statistically extendible to other datasets at a 5% significance level. The
authors concluded that their ANN model was an effective and robust approach for lesion
classification, performing better than the counterparts published earlier [47,48].
Horsch et al. explored three aspects of an LDA classifier that was based on automatic
segmentation of lesions and automatic extraction of lesion shape, margin, texture and posterior
acoustic behavior [50]. The study was conducted using a database of 400 cases with 94
malignancies, 124 complex cysts and 182 benign lesions. The reference standard was either
biopsy or aspiration. First, marginal benefit of adding a feature to the LDA model was
investigated. Second, the performance of the LDA model in distinguishing carcinomas from
different benign lesions was explored. The AUC values for the LDA model were 0.93 for
distinguishing carcinomas from complex cysts and 0.72 for differentiating fibrocystic disease
from carcinoma. Finally, eleven independent trials of training and testing were conducted to
validate the LDA model. Validation resulted in a mean AUC of 0.87 when computer-extracted
features from automatically delineated lesion margins were used. There was no statistically
significant difference between the best two- and four-feature classifiers; therefore, adding
features to the LDA model did not improve the performance.
Sahiner et al. investigated computer vision techniques to characterize breast tumors on 3D US
volumetric images [51]. The dataset was composed of masses from 102 women who underwent
either biopsy or fine-needle aspiration (56 had malignant masses). Automated mass
segmentation in 2D and 3D, as well as feature extraction followed by LDA, were implemented
to obtain malignancy scores. Stepwise feature selection was employed to reduce eight
morphologic and 72 texture features into a best-feature subset. An AUC of 0.87 was achieved
for the 2D-based classifier, while the AUC for the 3D-based classifier was 0.92. There was no
statistically significant difference between the two classifiers (p = 0.07). The AUC values of
the four radiologists fell in the range of 0.84 to 0.92. Comparing the performance of their model
to that of radiologists, the difference was not statistically significant (p = 0.05). However, the
partial AUC for their model was significantly higher than those of the three radiologists (p <
0.03, 0.02 and 0.001).
Drukker et al. used various feature segmentation and extraction schemes as inputs to a Bayesian
neural network (BNN) classifier with five hidden layers [52]. The purpose of the study was to
evaluate a CADx workstation in a realistic setting representative of clinical diagnostic breast
US practice. Benign or malignant lesions that were verified at biopsy or aspiration, as well as those determined through imaging characteristics on US scans, MR images and mammograms,
were used for the analysis. The authors included non-biopsied lesions in the dataset to make
the series consecutive, which more accurately reflects clinical practice. The inputs to the
network included lesion descriptors consisting of the depth:width ratio, radial gradient index,
posterior acoustic signature and autocorrelation texture feature. The output of the network
represented the probability of malignancy. The study was conducted on a patient population
of 508 (101 had breast cancer) with 1046 distinct abnormalities (157 cancerous lesions).
Comparing the current radiology practice with the CADx workstation, the CADx scheme
achieved an AUC of 0.90, corresponding to 100% sensitivity at 30% specificity, while
radiologists performed with 77% specificity for 100% sensitivity when only nonbiopsied
lesions were included. When only biopsy-proven lesions were analyzed, computerized lesion
characterization outperformed the radiologists.
In routine clinical practice, radiologists often combine the results from mammography and US,
if available, when making diagnostic decisions. Several studies demonstrated that CADx could
be useful in the differentiation of benign findings from malignant breast masses when
sonographic data are combined with corresponding mammographic data. Horsch et al.
evaluated and compared the performance of five radiologists with different expertise levels
and five imaging fellows with or without the help of a BNN [53]. The BNN model utilized a
computerized segmentation of the lesion. Mammographic features used as the input included
spiculation, lesion shape, margin sharpness, texture and gray level. Sonographic input features
included lesion shape, margin, texture and posterior acoustic behavior. All features were
automatically extracted by an image-processing algorithm. This retrospective study examined
a total of 359 (199 malignant) mammographic and 358 (67 malignant) sonographic images.
Additionally, 97 (39 malignant) multimodality cases (both mammogram and sonogram) were
used for testing purposes only. Biopsy was the reference standard. The performances of each
radiologist/imaging fellow or pair of observers were quantified by the AUC, sensitivity and
specificity metrics. Average AUC without BNN was 0.87 and with BNN was 0.92 (p < 0.001).
The sensitivities without and with BNN were 0.88 and 0.93, respectively (p = 0.005). There
was not a significant difference in specificities without and with BNN (0.66 vs 0.69, p = 0.20).
The authors concluded that the performance of the radiologists and imaging fellows increased
significantly with the help of the BNN model.
In another multimodality study, Sahiner et al. investigated the effect of a multimodal CADx
system (using mammography and US data) in discriminating between benign and malignant
lesions [54]. The dataset for the study consisted of 13 mammography features (nine
morphologic, three spiculation and one texture) and eight 3D US features (two morphologic
and six texture) that were extracted from 67 biopsy-proven masses (35 malignant). Ten
experienced readers first gave a malignancy score based on mammography only, then re-
evaluated based on mammography and US combined, and were finally allowed to change their
minds given the CADx system’s evaluation of the mass. The CADx system automatically
extracted the features, which were then fed into a multimodality classifier (using LDA) to give
a risk score. The results were compared using ROC curves, which suggested statistically
significant improvement (p = 0.05) when the CADx system was consulted (average AUC =
0.95) over readers’ assessment of combined mammography and US without the CADx
(average AUC = 0.93). Sahiner et al. concluded that a CADx system combining the features
from mammography and US may have the potential to improve radiologist’s diagnostic
decisions [54].
As discussed previously, a variety of sonographic features (texture, margin and shape) are used
to classify benign and malignant lesions. 2D/3D Doppler imaging provides additional
advantages in classification when compared with grayscale, by demonstrating breast lesion
vascularity. Chang et al. extracted features of tumor vascularity from 3D power Doppler US images of 221 lesions (110 benign) and devised an ANN to classify lesions [55]. The study
demonstrated that CADx, using 3D power Doppler imaging, can aid in the classification of
benign and malignant lesions.
In addition to the aforementioned studies, there are other works that developed and evaluated
CADx systems in differentiating between benign and malignant lesions. Joo et al. developed
an ANN that was demonstrated to have potential to increase the specificity of US
characterization of breast lesions [56]. Song et al. compared an LR and an ANN in the context
of differentiating between malignant and benign masses on breast sonograms from a small
dataset [57]. There was no statistically significant difference between the performances of the
two methods. Shen et al. investigated the statistical correlation between the computerized
sonographic features, as defined by BI-RADS, and the signs of malignancy [58]. Chen and
Hsiao evaluated US-based CADx systems by reviewing the methods used in classification
[59]. They suggested the inclusion of pathologically specific tissue-and hormone-related
features in future CADx systems. Gruszauskas et al. examined the effect of image selection
on the performance of a breast US CADx system and concluded that their automated breast
sonography classification scheme was reliable even with variation in user input [60]. Recently,
Cui et al. published a study focusing on the development of an automated method segmenting
and characterizing the breast masses on US images [61]. Their CADx system performed
similarly whether it used automated segmentation or an experienced radiologist’s
segmentation. In a recent study, Yap et al. designed a survey to evaluate the benefits of
computerized processing of US images in improving the readers’ performance of breast cancer
detection and classification [62]. The study demonstrated marginal improvements in
classification when computer-processed US images alongside the originals are used in
distinguishing benign from malignant lesions.


Thứ Ba, 11 tháng 11, 2014

O-MODE ULTRASOUND



O-mode ultrasound reduces artifacts
November 6, 2014 --
Sunday, November 30 | 12:05 p.m.-12:15 p.m. | SSA21-09 | Room S405AB
In this Sunday presentation, researchers will discuss how O-mode ultrasound is a promising technique for reducing artifacts typical with B-mode imaging.
Conventional ultrasound is limited by the need for beamforming, according to presenter Dr. Richard Barr, PhD, of Northeastern Ohio Medical University, and colleagues. O-mode ultrasound does not require a beamformer, offers constant lateral resolution, and decreases the shadowing that can arise from small pockets of gas. Barr's team conducted a pilot study to evaluate the use of O-mode imaging in a clinical setting.
The study included 10 patients who each received a complete abdominal ultrasound exam on a conventional ultrasound scanner as well as an O-mode device. A group of experienced ultrasound radiologists compared the images, ranking them on depth of penetration, shadowing artifacts, and overall image quality.
For all 10 cases, O-mode ultrasound visualized deep structures as well as or better than conventional B-mode ultrasound, the researchers found. In addition, O-mode ultrasound was able to visualize areas shadowed by ribs or small pockets of gas that B-mode could not.

O-mode Ultrasound, A New Novel Technique

Richard Gary Barr MD, PhD (Presenter): Consultant, Siemens AG Consultant, Koninklijke Philips NV Research Grant, Siemens AG Research Grant, SuperSonic Imagine Speakers Bureau, Koninklijke Philips NV Research Grant, Bracco Group Speakers Bureau, Siemens AG Consultant, Toshiba Corporation Research Grant, Esaote SpA , Alex Lomes PhD : Stockholder, Orcasonix Ltd , Mati Shirizly PhD : Shareholder, Orcasonix Ltd.

PURPOSE
Conventional US limitations and artifacts are related to the need for beam forming. O-mode introduces a novel approach that does not require a beam former, provides constant lateral resolution independent on depth of penetration and significantly decreases the shadowing from small pockets of gas allowing for new imaging windows to be utilized. The Doppler effect is evoked artificially by transmitting US signal from a "moving" transducer. Such transmitting scheme creates echo with different RF frequency shifts coming from different scatterers, which are located at the same depth, but at different lateral positions. This method provides exact lateral localization by exchanging the traditional focusing procedure (along line-of-sight) to signal processing of frequency-modulated signals coming from each depth, maintaining constant lateral resolution and improved
shadowing performance. This is a pilot study to evaluate O-mode in a clinical setting.

METHOD AND MATERIALS
10 patients participated in an IRB approved, HIPPA compliant study to evaluate the possible potential advantages of the O-mode imaging. Patients received a complete abdominal exam on a conventional ultrasound system (Esaote, My Lab Twice, Genoa, IT; Philips IU22, Bothell, Wa, or Siemens S3000, Mountain View, Ca) in addition to the Orcasonix O-mode system. Comparison of the images was performed by visual inspection by a board certified Radiologists with 20 years experience in ultrasound imaging. Images were scored as to depth of penetration, shadowing artifacts, and overall acceptability of image quality.

RESULTS
In all 10 cases O-mode was able to visualize deep structures equally or better than conventional B-mode. In areas of shadowing from ribs or small pockets of gas O-mode was able to visualize structures deep to the bone or gas which were able to be visualized on B-mode.

CONCLUSION
Preliminary clinical validation of O-mode imaging is promising. Additional post processing is required to improve O-mode image quality. O-mode imaging has several advantages over B-mode imaging including increased deep visualization, marked decreased shadowing and refractive artifacts. Without the need for a beam former smaller lighter ultrasound systems are possible.

CLINICAL RELEVANCE/APPLICATION
O-mode imaging can improve ultrasound imaging by significantly decreasing artifacts seen in B-mode imaging. The lack of a beam former allows for smaller ultrasound systems. 

ARTIFACTS in DIAGNOSTIC ULTRASOUND






Thứ Hai, 3 tháng 11, 2014

AAA: ULTRASOUND GUIDELINES

III. Indications/Contraindications

Indications for ultrasound of the abdominal aorta include but are not limited to:
A. Diagnostic Evaluation for an Abdominal Aortic
Aneurysm 
1. Palpable or pulsatile abdominal mass.
2. Unexplained lower back pain , flank pain, or abdominal pain.
3. Follow-up of a previously demonstrated abdominal aortic aneurysm.
4. Follow-up of patients with an abdominal aortic and/or iliac endoluminal stent graft.

B. Screening Evaluation for an Abdominal Aortic Aneurysm
1. Men 65 years or older.
2. Women 65 years or older with cardiovascular risk  factors.
3. Patients 50 years or older with a family history of aortic and/or peripheral vascular aneurysmal
disease.
4. Patients with a personal history of peripheral vascular aneurysmal disease.
Groups with additional risk include patients with a history of smoking, hypertension, and certain connective tissue diseases (eg, Marfan syndrome).
There are no absolute contraindications to ultrasound of the aorta. If aortic rupture or dissection is clinically suspected, ultrasound is usually not the examination of choice.

IV. Written Request for the Examination

The written or electronic request for an ultrasound examination should provide sufficient information  to allow for the appropriate performance and interpretation of the examination.

The request for the examination must be originated by a physician or other appropriately licensed health care provider or under their direction. The accompanying clinical information should be provided by a physician or other appropriate health care provider familiar with the
patient’s clinical situation and should be consistent with relevant legal and local health care facility requirements.

V. Specifications of the Examination
A. Diagnostic Examination
The examination includes the following, when feasible:
1. Abdominal aorta:
a. Longitudinal images (along the long axis of the vessel):
i. Proximal;
ii. Mid;
iii. Distal.

b. Transverse images (perpendicular to the long axis of the vessel):
i. Proximal (near diaphragm);
ii. Mid;
iii. Distal.
c. Measurements:
i. Measurements of the proximal, mid, and distal aorta should be obtained.
Measurements are taken at the greatest diameter of the aorta from outer edge to outer edge.
ii. If an aneurysm is present, the maximal size and location of the aneurysm should be documented and recorded. The relationship of the dilated segment to the renal arteries and to the aortic bifurcation should be determined if possible.
iii. A measurement of the length of the aneurysm is not necessary.
2. Common iliac arteries:
a. Longitudinal images of the proximal right and left common iliac arteries (along the long axis of the vessel).
b. Transverse images (perpendicular to the long axis of the vessel) of the proximal common iliac arteries just below the bifurcation.
c. Measurement of the widest visualized portion of each common iliac artery from outer edge to outer edge.
Color Doppler and/or spectral Doppler imaging with waveform analysis of the aorta and iliac arteries may provide additional information.
After endoluminal graft placement, color (or power) Doppler imaging and spectral Doppler imaging are required to document the presence or absence of endoleaks.
Interobserver measurements of an aortic aneurysm can vary by as much as 5 mm. This variation makes visual comparison with previous studies particularly important to determine whether a significant change in size has occurred.

B. Screening Examination for an Abdominal Aortic Aneurysm
1. Abdominal aorta:
a. Longitudinal images (along the long axis of the vessel):
i. Proximal;
ii. Mid;
iii. Distal.
b. Transverse images (perpendicular to the long axis of the vessel):
i. Proximal (near diaphragm);
ii. Mid;
iii. Distal.
C. Interpretation of the screening examination should include at least 3 categories:
1. Positive—Infrarenal abdominal aortic aneurysm greater than or equal to 3 cm in diameter or greater than or equal to 1.5 times the diameter of
the more proximal aorta.
The latter definition is particularly important in women.
2. Negative—No infrarenal abdominal aortic aneurysm.
3. Indeterminate—Aneurysmal status not defined because of nonvisualization or only partial visualization of the infrarenal abdominal aorta.
The report should also state whether the suprarenal aorta was seen and, if seen, should reflect whether it is normal.